Artis Zee and Zeego, model no. 10272462 AXIOM Artis zee/Zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Reported: December 12, 2018 Initiated: October 26, 2018 #Z-0558-2019

Product Description

Artis Zee and Zeego, model no. 10272462 AXIOM Artis zee/Zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Reason for Recall

This recall has been initiated due to a possible leakage of a hose in the cooling unit of the X-ray tube. The Artis zee system may lose coolant due to a possible leakage in a hose of the cooling unit in the X-ray tube. This will cause the X-ray tube not to be sufficiently cooled down. As a result, the system will display a message Tube hot, have a break . The X-ray will be blocked several minutes later and to prevent any damage to the machine the following message will be displayed NO XRAY, TUBE TOO HOT . This issue is very sporadic and may occur during an ongoing clinical procedure. Use of this product may have a potential impact on planned procedures as they may be delayed as well as in emergent cases, as clinical treatment may need to be terminated and transferred to a functioning system.

Details

Units Affected
64 (16 in US)
Distribution
US Distribution to states of: AR, CA, KS, KY, MD, MN, NM, OK, PA, RI, TX, and UT.
Location
Malvern, PA

Frequently Asked Questions

What product was recalled?
Artis Zee and Zeego, model no. 10272462 AXIOM Artis zee/Zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.. Recalled by Siemens Medical Solutions USA, Inc. Units affected: 64 (16 in US).
Why was this product recalled?
This recall has been initiated due to a possible leakage of a hose in the cooling unit of the X-ray tube. The Artis zee system may lose coolant due to a possible leakage in a hose of the cooling unit in the X-ray tube. This will cause the X-ray tube not to be sufficiently cooled down. As a result, the system will display a message Tube hot, have a break . The X-ray will be blocked several minutes later and to prevent any damage to the machine the following message will be displayed NO XRAY, TUBE TOO HOT . This issue is very sporadic and may occur during an ongoing clinical procedure. Use of this product may have a potential impact on planned procedures as they may be delayed as well as in emergent cases, as clinical treatment may need to be terminated and transferred to a functioning system.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 12, 2018. Severity: Moderate. Recall number: Z-0558-2019.