Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 30cm (reinforced), Material CA30L2

Reported: February 14, 2018 Initiated: August 7, 2017 #Z-0561-2018

Product Description

Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 30cm (reinforced), Material CA30L2

Reason for Recall

The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.

Details

Recalling Firm
Covidien LLC
Units Affected
N/A
Distribution
Nationwide including PR, Canada, China
Location
North Haven, CT

Frequently Asked Questions

What product was recalled?
Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 30cm (reinforced), Material CA30L2. Recalled by Covidien LLC. Units affected: N/A.
Why was this product recalled?
The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.