EMPOWR Locking Femoral Impactor , Part # 800-05-035, packaged in instrument tray FA EMP PAT TOOL KIT Part # 800-99-096
Product Description
EMPOWR Locking Femoral Impactor , Part # 800-05-035, packaged in instrument tray FA EMP PAT TOOL KIT Part # 800-99-096
Reason for Recall
The firm has received reports/complaints concerning the breakage of the tips of the locking femoral Impactor. Additionally, if the impactor is not seated correctly on the implant prior to impaction, this may cause off axis loading which can increase the risk of device failure. This could cause a delay in surgery. If the tip is not seen during surgery, long term risks may include an inflammatory response, infection, and/ or pain
Details
- Recalling Firm
- Encore Medical, Lp
- Units Affected
- 410 devices
- Distribution
- CO, WA, CA, UT, CA, TN, TX, MS, MO, AR, OK, LA, MN, IL, IN, KY, OH, MI, PA, SC, FL, PR, AL, GA, NM, AZ, NV, VA, RI, MD, ME PA, NY OUS: None
- Location
- Austin, TX
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11