VITEK 2 Antimicrobial Susceptibility Testing (AST) using one or more of these susceptibility products: Catalog # 410028, VITEK 2 AST-ST01 Catalog # 420915, VITEK 2 AST-ST02 Catalog # 421040, VITEK 2 AST-ST03 The VITEK 2 Streptococcus Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of S. pneumoniae, beta-hemolytic Streptococcus, and Viridans Streptococcus to antimicrobial agents when used as ins
Product Description
VITEK 2 Antimicrobial Susceptibility Testing (AST) using one or more of these susceptibility products: Catalog # 410028, VITEK 2 AST-ST01 Catalog # 420915, VITEK 2 AST-ST02 Catalog # 421040, VITEK 2 AST-ST03 The VITEK 2 Streptococcus Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of S. pneumoniae, beta-hemolytic Streptococcus, and Viridans Streptococcus to antimicrobial agents when used as instructed.
Reason for Recall
False resistant results for Streptococcus anginosus and Streptococcus constellatus strains were reported.
Details
- Recalling Firm
- bioMerieux, Inc.
- Units Affected
- 184,374
- Distribution
- Worldwide Distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI, WV, WY including Puerto Rico and to countries of:Algeria, Angola, Azerbaijan, Bahrain, Bangladesh, Belarus, Sarl, Benin, Bosnia-Herz, Cameroon, Chad, Cote d'Ivoire, Croatia, Dem. Rep. Congo, Egypt, , Georgia, Guam, Ethiopia, Iraq, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, UAB, Lithuania, Macedonia, Malta, Mongolia, Morocco, Mozambique, Niger, Oman, Pakistan, Palestine, Qatar, Romania, Saudi Arabia, Senegal, Serbia, Slovenia, Sri Lanka, Tunisia, Ukraine, Gabon, and Zimbabwe.
- Location
- Durham, NC
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11