Plum A+ Single Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 11971; b) list 11973 - software version 10.3; c) list 12391 - software version 11.3; d) list 20679 - Hospira MedNet Software e) list 20792 - Hospira MedNet Software; Product Usage: The Plum A+ Infusion System is a dual-line volumetric infusion pump that provides precise delivery of multiple i.v. therapies across the general spectrum of clinical care, inclu

Reported: January 9, 2013 Initiated: December 7, 2012 #Z-0624-2013

Product Description

Plum A+ Single Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 11971; b) list 11973 - software version 10.3; c) list 12391 - software version 11.3; d) list 20679 - Hospira MedNet Software e) list 20792 - Hospira MedNet Software; Product Usage: The Plum A+ Infusion System is a dual-line volumetric infusion pump that provides precise delivery of multiple i.v. therapies across the general spectrum of clinical care, including ICU/CCU, OR, PACU, ER, Med/Surg, Pediatrics, Outpatient clinics and Home care

Reason for Recall

The door roller assembly on the Plum A+/A+3 infuser family of infusers has the potential to break which can lead to possible unrestricted flow and/or over-delivery during the removal of the IV administration set's cassette from the pump.

Details

Recalling Firm
Hospira Inc.
Units Affected
297,833 units
Distribution
Worldwide Distribution -- USA, including Puerto Rico, and the countries of Austria, Australia, Argentina, Bahrain, Belgium, Brazil, Canada, Chile, Colombia, Denmark, Finland, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Zealand, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, United Arab Emirates, and Vietnam.
Location
Lake Forest, IL

Frequently Asked Questions

What product was recalled?
Plum A+ Single Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 11971; b) list 11973 - software version 10.3; c) list 12391 - software version 11.3; d) list 20679 - Hospira MedNet Software e) list 20792 - Hospira MedNet Software; Product Usage: The Plum A+ Infusion System is a dual-line volumetric infusion pump that provides precise delivery of multiple i.v. therapies across the general spectrum of clinical care, including ICU/CCU, OR, PACU, ER, Med/Surg, Pediatrics, Outpatient clinics and Home care. Recalled by Hospira Inc.. Units affected: 297,833 units.
Why was this product recalled?
The door roller assembly on the Plum A+/A+3 infuser family of infusers has the potential to break which can lead to possible unrestricted flow and/or over-delivery during the removal of the IV administration set's cassette from the pump.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 9, 2013. Severity: Moderate. Recall number: Z-0624-2013.