Plum A+3 Triple Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 12348 - software version 10.3; b) list 12618 - software version 11.3; c) list 20678 - Hospira MedNet software; Product Usage: The Plum A+3 is a dual-line triple channel volumetric infusion system that provides precise delivery of multiple i.v. therapies across the general spectrum of clinical care, including ICU/CCU, OR, PACU, ER, Med/Surg, Pediatrics, Out

Reported: January 9, 2013 Initiated: December 7, 2012 #Z-0625-2013

Product Description

Plum A+3 Triple Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 12348 - software version 10.3; b) list 12618 - software version 11.3; c) list 20678 - Hospira MedNet software; Product Usage: The Plum A+3 is a dual-line triple channel volumetric infusion system that provides precise delivery of multiple i.v. therapies across the general spectrum of clinical care, including ICU/CCU, OR, PACU, ER, Med/Surg, Pediatrics, Outpatient clinics and Home care

Reason for Recall

The door roller assembly on the Plum A+/A+3 infuser family of infusers has the potential to break which can lead to possible unrestricted flow and/or over-delivery during the removal of the IV administration set's cassette from the pump.

Details

Recalling Firm
Hospira Inc.
Units Affected
28,599 units
Distribution
Worldwide Distribution -- USA, including Puerto Rico, and the countries of Austria, Australia, Argentina, Bahrain, Belgium, Brazil, Canada, Chile, Colombia, Denmark, Finland, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Zealand, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, United Arab Emirates, and Vietnam.
Location
Lake Forest, IL

Frequently Asked Questions

What product was recalled?
Plum A+3 Triple Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 12348 - software version 10.3; b) list 12618 - software version 11.3; c) list 20678 - Hospira MedNet software; Product Usage: The Plum A+3 is a dual-line triple channel volumetric infusion system that provides precise delivery of multiple i.v. therapies across the general spectrum of clinical care, including ICU/CCU, OR, PACU, ER, Med/Surg, Pediatrics, Outpatient clinics and Home care. Recalled by Hospira Inc.. Units affected: 28,599 units.
Why was this product recalled?
The door roller assembly on the Plum A+/A+3 infuser family of infusers has the potential to break which can lead to possible unrestricted flow and/or over-delivery during the removal of the IV administration set's cassette from the pump.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 9, 2013. Severity: Moderate. Recall number: Z-0625-2013.