FDA Devices Moderate Class II Ongoing

Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10, Model Number: 722010. 2. Allura Xper FD10/10, Model Number: 722011. 3. Allura Xper FD20, Model Number: 722012. 4. Allura Xper FD20 Biplane, Model Number: 722013. 5. Allura Xper FD10 OR Table, Model Number: 722022. 6. Allura Xper FD20 OR Table, Model Number: 722023. 7. Allura Xper FD20 Biplane OR Table, Model Number: 722025.

Reported: December 18, 2024 Initiated: November 11, 2024 #Z-0676-2025

Product Description

Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10, Model Number: 722010. 2. Allura Xper FD10/10, Model Number: 722011. 3. Allura Xper FD20, Model Number: 722012. 4. Allura Xper FD20 Biplane, Model Number: 722013. 5. Allura Xper FD10 OR Table, Model Number: 722022. 6. Allura Xper FD20 OR Table, Model Number: 722023. 7. Allura Xper FD20 Biplane OR Table, Model Number: 722025.

Reason for Recall

Potential safety issues in Allura Xper R7.6- R8.1 systems where the Lifetime Extension (LTE) kit is not installed in a timely manner. Root cause is under investigation, but the recall reason is listed as for systems over 10 years of life with LTE kits are experiencing propeller motor clamping bolt and x-ray tube locking bolt breaks. If this happens, it may cause erratic C-arm movements and potential collision with patients or bystanders, or detection of a false collisions and subsequent blocked movement of the c-arm. Movement may also cause image noise and black area on the x-ray image, and possible need to expose the patient to additional unnecessary radiation during re-examination.

Details

Units Affected
2,782 units
Distribution
Nationwide distribution. International distribution to Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Oman, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen.
Location
Best

Frequently Asked Questions

What product was recalled?
Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10, Model Number: 722010. 2. Allura Xper FD10/10, Model Number: 722011. 3. Allura Xper FD20, Model Number: 722012. 4. Allura Xper FD20 Biplane, Model Number: 722013. 5. Allura Xper FD10 OR Table, Model Number: 722022. 6. Allura Xper FD20 OR Table, Model Number: 722023. 7. Allura Xper FD20 Biplane OR Table, Model Number: 722025.. Recalled by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. Units affected: 2,782 units.
Why was this product recalled?
Potential safety issues in Allura Xper R7.6- R8.1 systems where the Lifetime Extension (LTE) kit is not installed in a timely manner. Root cause is under investigation, but the recall reason is listed as for systems over 10 years of life with LTE kits are experiencing propeller motor clamping bolt and x-ray tube locking bolt breaks. If this happens, it may cause erratic C-arm movements and potential collision with patients or bystanders, or detection of a false collisions and subsequent blocked movement of the c-arm. Movement may also cause image noise and black area on the x-ray image, and possible need to expose the patient to additional unnecessary radiation during re-examination.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 18, 2024. Severity: Moderate. Recall number: Z-0676-2025.