PlainRecalls
FDA Devices Moderate Class II Ongoing

42 IN (107 cm) Appx 5.0 ml, 20 Drop Admin Set, Spiros, REF: 20131-01; 31" (79 cm) Admin Set w/20 Drop Integrated Clave Drip Chamber, Spiros, Bag Hanger Drop-in Red Cap, 4 Units, REF: CH3330-C; 60 IN (152 cm) Appx 1.1 ml, Smallbore Set, MicroClave Clear, Spiros, REF: CH3353; 8" (20 cm) Appx 0.43 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, 2 Clamps, Spiros, REF: CH3367; 7" (18 cm) Appx 0.40 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, Spiros, 2 Clamps, REF: CH3379; 16" (41 cm) Appx 4

Reported: December 30, 2020 Initiated: August 31, 2020 #Z-0699-2021

Product Description

42 IN (107 cm) Appx 5.0 ml, 20 Drop Admin Set, Spiros, REF: 20131-01; 31" (79 cm) Admin Set w/20 Drop Integrated Clave Drip Chamber, Spiros, Bag Hanger Drop-in Red Cap, 4 Units, REF: CH3330-C; 60 IN (152 cm) Appx 1.1 ml, Smallbore Set, MicroClave Clear, Spiros, REF: CH3353; 8" (20 cm) Appx 0.43 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, 2 Clamps, Spiros, REF: CH3367; 7" (18 cm) Appx 0.40 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, Spiros, 2 Clamps, REF: CH3379; 16" (41 cm) Appx 4.4 ml, Ext Set w/MicroClave, 0.2 Micron Filter, Clave, Spiros, Clamp, REF: CH3381; 31" (79 cm) Appx 3.3 ml, Admin Set w/20 Drop Integrated Clave Drip Chamber, Spiros w/Drop-in Red Cap, 2 Hanger, REF: CH3386; 40" (102 cm) Appx 7.3 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, 0.2 Micron Filter, Spiros, Hanger, REF: CH3401; 60" (152 cm) Appx 0.62 ml, Smallbore Ext Set w/MicroClave, Clamp, Spiros w/Red Cap, REF: CH3405; 75" (191 cm) Appx 9.1 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Clave, Spiros w/Red Cap, REF: CH3453; 73" (185 cm) Appx 2.1 ml, Smallbore Ext Set w/MicroClave Clear, Spiros, Clamp, REF: CH3485; 42" (107 cm) Appx 4.8 ml, 15 Drop Admin Set w/Spiros w/Red Cap, Bag Hanger, Drop-In Priming Cap, REF: CH3517; 8" (20 cm) Appx 0.49 ml, Smallbore Bifuse Ext Set w/MicroClave Clear, 2 Spiros, 3 Clamps, Priming Cap, REF: CH3523; 74" (188 cm) Smallbore Ext Set w/MicroClave, 0.2 Micron Low Protein Binding Filter, Spiros w/Red Cap, Clamp, REF: CH3533; 74 IN (188 cm) Appx 0.92 ml, Smallbore Set, MicroClave, 1.2 Micron Fil, Spiros, Red Cap, REF: CH3538; 18" (46 cm) Appx 5.1 ml, Ext Set w/0.2 Micron Filter, Clave, Spiros, Clamp, REF: CH3556; 60" (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave Clear, 0.2 Micron Low Protein Binding Filter, Spiros w/Red Cap, Clamp, REF: CH3603; 4-Way Clave Stopcock w/2 Spiros, REF: B4194; Spiros, Closed Male Connector w/Red Cap (10 Units), REF: Z7093; 47" (119 cm) 150 ml Burette Set w/MicroClave, 20 Drop In-Line Drip Chamber w/Ball Shut Off, Spiros, REF: Z7182

Reason for Recall

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

Details

Recalling Firm
ICU Medical, Inc.
Units Affected
19850
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.
Location
San Clemente, CA

Frequently Asked Questions

What product was recalled?
42 IN (107 cm) Appx 5.0 ml, 20 Drop Admin Set, Spiros, REF: 20131-01; 31" (79 cm) Admin Set w/20 Drop Integrated Clave Drip Chamber, Spiros, Bag Hanger Drop-in Red Cap, 4 Units, REF: CH3330-C; 60 IN (152 cm) Appx 1.1 ml, Smallbore Set, MicroClave Clear, Spiros, REF: CH3353; 8" (20 cm) Appx 0.43 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, 2 Clamps, Spiros, REF: CH3367; 7" (18 cm) Appx 0.40 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, Spiros, 2 Clamps, REF: CH3379; 16" (41 cm) Appx 4.4 ml, Ext Set w/MicroClave, 0.2 Micron Filter, Clave, Spiros, Clamp, REF: CH3381; 31" (79 cm) Appx 3.3 ml, Admin Set w/20 Drop Integrated Clave Drip Chamber, Spiros w/Drop-in Red Cap, 2 Hanger, REF: CH3386; 40" (102 cm) Appx 7.3 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, 0.2 Micron Filter, Spiros, Hanger, REF: CH3401; 60" (152 cm) Appx 0.62 ml, Smallbore Ext Set w/MicroClave, Clamp, Spiros w/Red Cap, REF: CH3405; 75" (191 cm) Appx 9.1 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Clave, Spiros w/Red Cap, REF: CH3453; 73" (185 cm) Appx 2.1 ml, Smallbore Ext Set w/MicroClave Clear, Spiros, Clamp, REF: CH3485; 42" (107 cm) Appx 4.8 ml, 15 Drop Admin Set w/Spiros w/Red Cap, Bag Hanger, Drop-In Priming Cap, REF: CH3517; 8" (20 cm) Appx 0.49 ml, Smallbore Bifuse Ext Set w/MicroClave Clear, 2 Spiros, 3 Clamps, Priming Cap, REF: CH3523; 74" (188 cm) Smallbore Ext Set w/MicroClave, 0.2 Micron Low Protein Binding Filter, Spiros w/Red Cap, Clamp, REF: CH3533; 74 IN (188 cm) Appx 0.92 ml, Smallbore Set, MicroClave, 1.2 Micron Fil, Spiros, Red Cap, REF: CH3538; 18" (46 cm) Appx 5.1 ml, Ext Set w/0.2 Micron Filter, Clave, Spiros, Clamp, REF: CH3556; 60" (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave Clear, 0.2 Micron Low Protein Binding Filter, Spiros w/Red Cap, Clamp, REF: CH3603; 4-Way Clave Stopcock w/2 Spiros, REF: B4194; Spiros, Closed Male Connector w/Red Cap (10 Units), REF: Z7093; 47" (119 cm) 150 ml Burette Set w/MicroClave, 20 Drop In-Line Drip Chamber w/Ball Shut Off, Spiros, REF: Z7182. Recalled by ICU Medical, Inc.. Units affected: 19850.
Why was this product recalled?
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 30, 2020. Severity: Moderate. Recall number: Z-0699-2021.