XScribe CP Cardiac Stress Testing System, version 6, with below product names and codes: a) X-SCRIBE CPI KIT WIRELESS USB 608 MHz, 41000-030-50; b) X-SCRIBE CPI KIT WIRELESS USB 2500 MHz, 41000-030-51; c) X-SCRIBE CPI KIT DIRECT CONNECT USB, 41000-030-52; d) X-SCRIBE CPI KIT QUICK CONNECT USB, 41000-030-53; e) XSCRIBE CP KIT WIRELESS, 41000-030-60; f) XSCRIBE CP KIT WIRED, 41000-030-62; g) XSCRIBE CP SW KIT, 41000-030-64

Reported: January 17, 2024 Initiated: December 8, 2023 #Z-0707-2024

Product Description

XScribe CP Cardiac Stress Testing System, version 6, with below product names and codes: a) X-SCRIBE CPI KIT WIRELESS USB 608 MHz, 41000-030-50; b) X-SCRIBE CPI KIT WIRELESS USB 2500 MHz, 41000-030-51; c) X-SCRIBE CPI KIT DIRECT CONNECT USB, 41000-030-52; d) X-SCRIBE CPI KIT QUICK CONNECT USB, 41000-030-53; e) XSCRIBE CP KIT WIRELESS, 41000-030-60; f) XSCRIBE CP KIT WIRED, 41000-030-62; g) XSCRIBE CP SW KIT, 41000-030-64

Reason for Recall

Potential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) is enabled.

Details

Units Affected
1675 units
Distribution
United States (including Puerto Rico), Canada, Australia, Bangladesh, Brunei, Chia, Hong Kong, India, Indonesia, Korea, Malaysia, New Zealand, Philippines, Singapore, Sri Lanka, Thailand, Algeria, Angola, Austria, Bahrain, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Hungary, Iceland, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libya, Morocco, Namibia, Netherlands, Norway, Oman, Pakistan, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, South Africa, Spain, Switzerland, Tunisia, Turkey, United Arab Emirates, United Kingdom, Argentina, Bahamas, Bolivia, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, Guatemala, Honduras, Mexico, Panama, Paraguay, Peru, and Trinidad and Tobago.
Location
Deerfield, IL

Frequently Asked Questions

What product was recalled?
XScribe CP Cardiac Stress Testing System, version 6, with below product names and codes: a) X-SCRIBE CPI KIT WIRELESS USB 608 MHz, 41000-030-50; b) X-SCRIBE CPI KIT WIRELESS USB 2500 MHz, 41000-030-51; c) X-SCRIBE CPI KIT DIRECT CONNECT USB, 41000-030-52; d) X-SCRIBE CPI KIT QUICK CONNECT USB, 41000-030-53; e) XSCRIBE CP KIT WIRELESS, 41000-030-60; f) XSCRIBE CP KIT WIRED, 41000-030-62; g) XSCRIBE CP SW KIT, 41000-030-64. Recalled by Baxter Healthcare Corporation. Units affected: 1675 units.
Why was this product recalled?
Potential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) is enabled.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 17, 2024. Severity: Moderate. Recall number: Z-0707-2024.