Reported: January 22, 2014 Initiated: December 10, 2013 #Z-0714-2014

Product Description

Draeger Infinity Acute Care System Monitoring Solution

Reason for Recall

After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.

Details

Units Affected
439
Distribution
US Distribution including the states of: MA, NH, IN, CA, TX, NY, OK, and IL.
Location
Telford, PA