Philips Allura Xper, UNIQ(Allura Xper system-China) Interventional Fluoroscopic X-ray System products with a Laird chiller WITHOUT an extended drip tray: Model Number: 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025, 722026, 722027, 722029, 722033, 722034, 722036, 722038, 722039. Allura Xper F010 {C+F). 722003 Allura Xper F010/10 722005 Allura Xper F020 722006 Allura Xper F020/10 and F020/20 722008 Allura Xpe

Reported: March 16, 2022 Initiated: November 24, 2021 #Z-0724-2022

Product Description

Philips Allura Xper, UNIQ(Allura Xper system-China) Interventional Fluoroscopic X-ray System products with a Laird chiller WITHOUT an extended drip tray: Model Number: 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025, 722026, 722027, 722029, 722033, 722034, 722036, 722038, 722039. Allura Xper F010 {C+F). 722003 Allura Xper F010/10 722005 Allura Xper F020 722006 Allura Xper F020/10 and F020/20 722008 Allura Xper F010 722010 Altura Xper F010/10 722011 Allura Xper F020 722012 Allura Xper FD20 biplane 722013 Allura Xper FDlO OR Table 722014 Allura Xper FD20 OR Table 722015 Allura Xper FDl0/10 OR Table 722019 Allura Xper FD20 Biplane OR Table 722020 Allura Xper FD10 OR Table 722022 Allura Xper FD20 OR Table 722023 Allura Xper FD10/10 OR Table 722024 Altura Xper FD20 Biplane OR Table 722025 Allura Xper FD10 722026 Allura Xper FD10/10 722027 Altura Xper FD20/10 biplane 722029 Allura Xper FD10 OR Table 722033 Allura Xper FD10/10 OR Table 722034 Allura Xper FD20/10 OR Table 722036 Allure Xper F020/20 722038 Altura Xper FD20/20 biplane OR Table 722039

Reason for Recall

Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.

Details

Recalling Firm
Philips North America Llc
Units Affected
USA: 63 systems R.O.W.: 127 systems
Distribution
US Nationwide Foreign: Argentina Austria Bangladesh Brazil Chile China Denmark Dominican Republic France Germany Greece India Indonesia Iran Israel Italy Japan Korea, Republic of Malaysia Mexico Netherlands Philippines Russian Federation Saudi Arabia South Africa Spain Sri Lanka Switzerland Taiwan Thailand Turkey United Kingdom Virgin Islands (U.S.)
Location
Cambridge, MA

Frequently Asked Questions

What product was recalled?
Philips Allura Xper, UNIQ(Allura Xper system-China) Interventional Fluoroscopic X-ray System products with a Laird chiller WITHOUT an extended drip tray: Model Number: 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025, 722026, 722027, 722029, 722033, 722034, 722036, 722038, 722039. Allura Xper F010 {C+F). 722003 Allura Xper F010/10 722005 Allura Xper F020 722006 Allura Xper F020/10 and F020/20 722008 Allura Xper F010 722010 Altura Xper F010/10 722011 Allura Xper F020 722012 Allura Xper FD20 biplane 722013 Allura Xper FDlO OR Table 722014 Allura Xper FD20 OR Table 722015 Allura Xper FDl0/10 OR Table 722019 Allura Xper FD20 Biplane OR Table 722020 Allura Xper FD10 OR Table 722022 Allura Xper FD20 OR Table 722023 Allura Xper FD10/10 OR Table 722024 Altura Xper FD20 Biplane OR Table 722025 Allura Xper FD10 722026 Allura Xper FD10/10 722027 Altura Xper FD20/10 biplane 722029 Allura Xper FD10 OR Table 722033 Allura Xper FD10/10 OR Table 722034 Allura Xper FD20/10 OR Table 722036 Allure Xper F020/20 722038 Altura Xper FD20/20 biplane OR Table 722039. Recalled by Philips North America Llc. Units affected: USA: 63 systems R.O.W.: 127 systems.
Why was this product recalled?
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
Which agency issued this recall?
This recall was issued by the FDA Devices on March 16, 2022. Severity: Moderate. Recall number: Z-0724-2022.