PlainRecalls
FDA Devices Critical Class I Ongoing

Medline Fluid Delivery Set with Drip Chamber: 1) MTO,135,YELLOW,MICRO,SPK-MAIMONIDES, Item Number: 60130527; 2) MTO,ANGIO,DOUBLE,SPIKE,KIT,KENNEDY HLTH, Item Number: 65111341; 3) MTO,ANGIO,SINGLE,SPIKE,KIT, Item Number: 65232486; 4) MTO,ANGIO,SINGLE,SPIKE,KIT,GOOD,SAM, Item Number: 65071193; 5) MTO,NEURO,FLUSH,SPIKE,KIT,BLUE,FLORIDA, Item Number: 600604422; 6) MTO,NEURO,KIT,-,JERSEY,SHORE PG, Item Number: 60100045; 7) MTO,RADIOLOGY,KIT,HAMOT MED CEN, Item Number: 650801715; 8) MTS,135

Reported: January 22, 2025 Initiated: December 16, 2024 #Z-0765-2025

Product Description

Medline Fluid Delivery Set with Drip Chamber: 1) MTO,135,YELLOW,MICRO,SPK-MAIMONIDES, Item Number: 60130527; 2) MTO,ANGIO,DOUBLE,SPIKE,KIT,KENNEDY HLTH, Item Number: 65111341; 3) MTO,ANGIO,SINGLE,SPIKE,KIT, Item Number: 65232486; 4) MTO,ANGIO,SINGLE,SPIKE,KIT,GOOD,SAM, Item Number: 65071193; 5) MTO,NEURO,FLUSH,SPIKE,KIT,BLUE,FLORIDA, Item Number: 600604422; 6) MTO,NEURO,KIT,-,JERSEY,SHORE PG, Item Number: 60100045; 7) MTO,RADIOLOGY,KIT,HAMOT MED CEN, Item Number: 650801715; 8) MTS,135,CDS,IV,TUBE-JEFFERSON UNIV., Item Number: 60232481; 9) MTS,135,YELLOW,MICRO,SPIKE,LUTHERAN MED, Item Number: 60120583; 10) MTS,ANGIO,NEURO,KIT,NORTH SHORE UNIV, Item Number: 601403920; 11) MTS,CUST. FILTER-NEWYORK UNIVERSITY, Item Number: 62140602; 12) MTS,INTERV,NEURO,KIT-WESTCHESTER,MC, Item Number: 602301614; 13) MTS,NEU,NORTH SHORE HOSPITAL, Item Number: 6010771601; 14) MTS,NEURO,KIT, Item Number: 600602914; 15) MTS,NEURO,KIT,2,NEWENGLAND MED CTR, Item Number: 601400111; 16) MTS,NEURO,KIT,YLW&RED,FLORIDA HOSP, Item Number: 600604423; 17) MTS,NEURO,KIT-JOHN,F,KENNEDY MED CENTER, Item Number: 60110552; 18) MTS,NEURO,KIT-OVERLOOK HOSPITAL PG, Item Number: 60150035; 19) MTS,NEURO,SPIKE,KIT,CAPITAL HLTHSYS-FULD, Item Number: 60182723; 20) MTS,NEURO-RADIOLOGY,KIT-YALE-NEWHAVEN, Item Number: 602500214; 21) MTS,STROKE/EMBO,KIT,NORTH SHORE UNIV, Item Number: 601403921; 22) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601305210; 23) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601400110;

Reason for Recall

Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.

Details

Units Affected
3930 units
Distribution
US Nationwide Distribution
Location
Northfield, IL

Frequently Asked Questions

What product was recalled?
Medline Fluid Delivery Set with Drip Chamber: 1) MTO,135,YELLOW,MICRO,SPK-MAIMONIDES, Item Number: 60130527; 2) MTO,ANGIO,DOUBLE,SPIKE,KIT,KENNEDY HLTH, Item Number: 65111341; 3) MTO,ANGIO,SINGLE,SPIKE,KIT, Item Number: 65232486; 4) MTO,ANGIO,SINGLE,SPIKE,KIT,GOOD,SAM, Item Number: 65071193; 5) MTO,NEURO,FLUSH,SPIKE,KIT,BLUE,FLORIDA, Item Number: 600604422; 6) MTO,NEURO,KIT,-,JERSEY,SHORE PG, Item Number: 60100045; 7) MTO,RADIOLOGY,KIT,HAMOT MED CEN, Item Number: 650801715; 8) MTS,135,CDS,IV,TUBE-JEFFERSON UNIV., Item Number: 60232481; 9) MTS,135,YELLOW,MICRO,SPIKE,LUTHERAN MED, Item Number: 60120583; 10) MTS,ANGIO,NEURO,KIT,NORTH SHORE UNIV, Item Number: 601403920; 11) MTS,CUST. FILTER-NEWYORK UNIVERSITY, Item Number: 62140602; 12) MTS,INTERV,NEURO,KIT-WESTCHESTER,MC, Item Number: 602301614; 13) MTS,NEU,NORTH SHORE HOSPITAL, Item Number: 6010771601; 14) MTS,NEURO,KIT, Item Number: 600602914; 15) MTS,NEURO,KIT,2,NEWENGLAND MED CTR, Item Number: 601400111; 16) MTS,NEURO,KIT,YLW&RED,FLORIDA HOSP, Item Number: 600604423; 17) MTS,NEURO,KIT-JOHN,F,KENNEDY MED CENTER, Item Number: 60110552; 18) MTS,NEURO,KIT-OVERLOOK HOSPITAL PG, Item Number: 60150035; 19) MTS,NEURO,SPIKE,KIT,CAPITAL HLTHSYS-FULD, Item Number: 60182723; 20) MTS,NEURO-RADIOLOGY,KIT-YALE-NEWHAVEN, Item Number: 602500214; 21) MTS,STROKE/EMBO,KIT,NORTH SHORE UNIV, Item Number: 601403921; 22) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601305210; 23) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601400110;. Recalled by MEDLINE INDUSTRIES, LP - Northfield. Units affected: 3930 units.
Why was this product recalled?
Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 22, 2025. Severity: Critical. Recall number: Z-0765-2025.