PlainRecalls
FDA Devices Moderate Class II Ongoing

Access Intact PTH assay, a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma using the Access Immunoassay Systems, Catalog Number A16972

Reported: January 15, 2025 Initiated: November 26, 2024 #Z-0803-2025

Product Description

Access Intact PTH assay, a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma using the Access Immunoassay Systems, Catalog Number A16972

Reason for Recall

Beckman Coulter has identified that some Access PTH reagent packs from lot 339071 may be over- or under-filled in well 0. The dispensed material in well 0 could be above or below specification, risking sealing issues, contamination, instrument errors, delays, or inaccurate results with a negative bias and increase in imprecision. The issue was confirmed on 10 April 2024 by Beckman Coulter through an internal non-conformance (NC-INT-66585). The well 0 scale on the filling line began malfunctioning on the fill immediately before PTH. The malfunction was not noticed until the majority of PTH was filled.

Details

Recalling Firm
Beckman Coulter, Inc.
Units Affected
362 units
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Location
Chaska, MN

Frequently Asked Questions

What product was recalled?
Access Intact PTH assay, a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma using the Access Immunoassay Systems, Catalog Number A16972. Recalled by Beckman Coulter, Inc.. Units affected: 362 units.
Why was this product recalled?
Beckman Coulter has identified that some Access PTH reagent packs from lot 339071 may be over- or under-filled in well 0. The dispensed material in well 0 could be above or below specification, risking sealing issues, contamination, instrument errors, delays, or inaccurate results with a negative bias and increase in imprecision. The issue was confirmed on 10 April 2024 by Beckman Coulter through an internal non-conformance (NC-INT-66585). The well 0 scale on the filling line began malfunctioning on the fill immediately before PTH. The malfunction was not noticed until the majority of PTH was filled.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 15, 2025. Severity: Moderate. Recall number: Z-0803-2025.