GE Healthcare, TRAM MODULE, High-acuity patient monitoring. Product Usage: The TRAM Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. K900540: TRAM modules are intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The intended use of TRAM modules is to
Product Description
GE Healthcare, TRAM MODULE, High-acuity patient monitoring. Product Usage: The TRAM Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. K900540: TRAM modules are intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The intended use of TRAM modules is to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. The TRAM module functions as a transport monitor during transport, and in the bedside monitor during ICU or OR monitoring. Physiological parameter data includes ECG, invasive blood pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature, and respiration. TRAM modules acquire, process and store information regarding these parameters. TRAM modules incorporate different monitoring capabilities based on their configuration. K011000: The TRAM 2001 Module is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The TRAM 2001 Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric, and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. During non-transport monitoring, the TRAM 2001 Module functions in the bedside monitoring system. Physiological parameter data includes ECG, invasive blood pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature, and respiration. The TRAM 2001 Module acquires, processes and stores information regarding these parameters. The device is intended for use in a professional medical facility, such as hospital, clinic, surgical center or doctors office. The TRAM System can be used in multiple areas such as operating room (OR), post anesthesia recovery (PARR), critical care, surgical intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal intensive care. The device is intended for use as part of a transport monitoring system for intra-hospital transport. K053121: The TRAM Module is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in its proper use. The TRAM Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. During non-transport monitoring, the TRAM Module functions in the bedside monitoring system. Physiological parameter data includes ECG, invasive blood pressure, non-invasive blood pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature and respiration. The TRAM Module acquires, processes and stores information regarding these parameters. The device is intended for use in a professional medical facility, such as hospital, clinic, surgical center or doctors office. The TRAM System can be used in multiple areas such as operating room (OR), post anesthesia recovery (PARR), critical care, surgical intensive care, pediatric intensive care, or neonatal intensive care. The device is intended for use as part of a transport monitoring system for intra-hospital transport.
Reason for Recall
Intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG cable connector on the DASH, PDM, and/or Tram Module. May result in reduced ECG, Respiration and arrhythmia detection performance.
Details
- Recalling Firm
- GE Healthcare, LLC
- Units Affected
- 1,948 (1,834 US; 114 OUS)
- Distribution
- Worldwide Distribution: US (nationwide) including DC and PR except SD, VT and WY; and Internationally to: ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, EGYPT, GERMANY, UNITED KINGDOM, IRELAND, IRAQ, ISRAEL, JAMAICA, JAPAN, JORDAN, KOREA, KUWAIT, LEBANON, LITHUANIA, MARITIUS, MALDOVA, NEW ZEALAND, NIGERIA, PAKISTAN, PANAMA, PARAGUAY, PERU, POLAND, QATAR, RUSSIA, SLOVAKIA, SOUTH AFRICA, SWEDEN, SWITZERLAND, TUNISIA, UNITED ARAB EMIRATES, VENEZUELA, CHILE, CANADA, and BERMUDA.
- Location
- Waukesha, WI
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11