PlainRecalls
FDA Devices Moderate Class II Terminated

BIOMET Sports Medicine JUGGERLOC SLOTTED REAMER: [Item # 110010371, 6MM], [Item #110010372, 7MM], [Item # 110010373, 8MM], [Item #110018275, 5MM]. Devices are packaged in a vacuum sealed nylon pouch, which is placed in a carton with the appropriate IFU and patient record labels. Identifying labels on the pouch and the carton.

Reported: March 30, 2022 Initiated: February 14, 2022 #Z-0809-2022

Product Description

BIOMET Sports Medicine JUGGERLOC SLOTTED REAMER: [Item # 110010371, 6MM], [Item #110010372, 7MM], [Item # 110010373, 8MM], [Item #110018275, 5MM]. Devices are packaged in a vacuum sealed nylon pouch, which is placed in a carton with the appropriate IFU and patient record labels. Identifying labels on the pouch and the carton.

Reason for Recall

Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile barrier integrity. Use of product poses risk of infection, leading to surgical intervention, and/or extension of surgery to find another part.

Details

Recalling Firm
Biomet, Inc.
Units Affected
702
Distribution
Worldwide: US (28 States AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KS, KY, LA, MD, MI, MS, NC, NE, NM, NY, OH, OK, OR, SC, TX, UT, WA, WY), Australia, Japan, and Netherlands
Location
Warsaw, IN

Frequently Asked Questions

What product was recalled?
BIOMET Sports Medicine JUGGERLOC SLOTTED REAMER: [Item # 110010371, 6MM], [Item #110010372, 7MM], [Item # 110010373, 8MM], [Item #110018275, 5MM]. Devices are packaged in a vacuum sealed nylon pouch, which is placed in a carton with the appropriate IFU and patient record labels. Identifying labels on the pouch and the carton.. Recalled by Biomet, Inc.. Units affected: 702.
Why was this product recalled?
Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile barrier integrity. Use of product poses risk of infection, leading to surgical intervention, and/or extension of surgery to find another part.
Which agency issued this recall?
This recall was issued by the FDA Devices on March 30, 2022. Severity: Moderate. Recall number: Z-0809-2022.