PlainRecalls
FDA Devices Moderate Class II Terminated

EPIQ 7 Ultrasound System, EPIQ 7 systems with software versions 1.0.x or 1.1.x with QLAB a2DQ and/or aCMQ plug-ins are affected. Model: EPIQ 7G, EPIC 7C, EPIQ 7W; Catalog Number: 795200 / 795201 and Part Number: 989605386721. Diagnostic Ultrasound Imaging.

Reported: December 24, 2014 Initiated: November 5, 2014 #Z-0817-2015

Product Description

EPIQ 7 Ultrasound System, EPIQ 7 systems with software versions 1.0.x or 1.1.x with QLAB a2DQ and/or aCMQ plug-ins are affected. Model: EPIQ 7G, EPIC 7C, EPIQ 7W; Catalog Number: 795200 / 795201 and Part Number: 989605386721. Diagnostic Ultrasound Imaging.

Reason for Recall

When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left Ventricular Ejection Fraction (EF) may be higher, than the ESV and EF calculated by manual tracing without the use of QLAB. An incorrect EF calculation could lead to misdiagnosis and/or delayed or incorrect therapy.

Details

Recalling Firm
Philips Ultrasound, Inc.
Units Affected
701 units (237 in the US and 464 outside US)
Distribution
Worldwide distribution (for FCO 83000190): US, including the states of AK, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NJ, NY, OH, OR, PA, SC, SD, TN, TX, VA, WA, and WI; and the countries of United Arab Emirates, Austria, Australia, Belgium, Canada, Switzerland, Colombia, Germany, Ecuador, Estonia, Spain, Finland, France, Great Britain, Hong Kong, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, Norway, New Zealand, Philippines, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Turkey, Taiwan, and Vietnam. ********************************************************************************** Worldwide Distribution -- US, including the states of KY, TN, ME, CA, PA, NC, CT, NY, SC, IL, IN, and WI; and the countries of Argentina, Spain, India, Italy, Austria, Germany, Belgium, Canada, Finland, Switzerland, Greece, Malaysia, Sweden, Australia, Netherlands, Mexico, Peru, Taiwan, Japan, France, and Norway.
Location
Bothell, WA

Frequently Asked Questions

What product was recalled?
EPIQ 7 Ultrasound System, EPIQ 7 systems with software versions 1.0.x or 1.1.x with QLAB a2DQ and/or aCMQ plug-ins are affected. Model: EPIQ 7G, EPIC 7C, EPIQ 7W; Catalog Number: 795200 / 795201 and Part Number: 989605386721. Diagnostic Ultrasound Imaging.. Recalled by Philips Ultrasound, Inc.. Units affected: 701 units (237 in the US and 464 outside US).
Why was this product recalled?
When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left Ventricular Ejection Fraction (EF) may be higher, than the ESV and EF calculated by manual tracing without the use of QLAB. An incorrect EF calculation could lead to misdiagnosis and/or delayed or incorrect therapy.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 24, 2014. Severity: Moderate. Recall number: Z-0817-2015.