PlainRecalls
FDA Devices Moderate Class II Terminated

Centricity Universal Viewer 6.0, a device that displays medical images and data from various imaging sources, and from other healthcare information sources.

Reported: January 29, 2020 Initiated: December 16, 2019 #Z-0819-2020

Product Description

Centricity Universal Viewer 6.0, a device that displays medical images and data from various imaging sources, and from other healthcare information sources.

Reason for Recall

Centricity Universal Viewer measurements saved into a DICOM Grayscale Presentation State are incorrect in subsequent views for exams containing series with different pixel sizes and may lead to a potential misdiagnosis.

Details

Recalling Firm
GE Healthcare, LLC
Units Affected
205
Distribution
Distributed nationwide; Internationally distributed to Mexico and 24 additional countries.
Location
Waukesha, WI

Frequently Asked Questions

What product was recalled?
Centricity Universal Viewer 6.0, a device that displays medical images and data from various imaging sources, and from other healthcare information sources.. Recalled by GE Healthcare, LLC. Units affected: 205.
Why was this product recalled?
Centricity Universal Viewer measurements saved into a DICOM Grayscale Presentation State are incorrect in subsequent views for exams containing series with different pixel sizes and may lead to a potential misdiagnosis.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 29, 2020. Severity: Moderate. Recall number: Z-0819-2020.