Synthes VA Implant Rack for the Compact Distal Radius System Tray/Surgical/Instrument

Reported: December 28, 2016 Initiated: November 29, 2016 #Z-0867-2017

Product Description

Synthes VA Implant Rack for the Compact Distal Radius System Tray/Surgical/Instrument

Reason for Recall

The screw rack plate was assembled incorrectly which could cause the screws in the screw rack to sit low and touch the surface beneath the screw rack.

Details

Units Affected
31 units
Distribution
Nationwide Distribution
Location
West Chester, PA

Frequently Asked Questions

What product was recalled?
Synthes VA Implant Rack for the Compact Distal Radius System Tray/Surgical/Instrument. Recalled by Synthes (USA) Products LLC. Units affected: 31 units.
Why was this product recalled?
The screw rack plate was assembled incorrectly which could cause the screws in the screw rack to sit low and touch the surface beneath the screw rack.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 28, 2016. Severity: Moderate. Recall number: Z-0867-2017.