FDA Devices Moderate Class II Ongoing

Philips Allura (Model Numbers 722010, 722012, 722026, 722027, 722028, 722029, 722035) and Philips Azurion (Model Numbers 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228) running StentBoost Live (Model Number 459801781661) Software Version Number: R2.0

Reported: April 13, 2022 Initiated: February 2, 2022 #Z-0874-2022

Product Description

Philips Allura (Model Numbers 722010, 722012, 722026, 722027, 722028, 722029, 722035) and Philips Azurion (Model Numbers 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228) running StentBoost Live (Model Number 459801781661) Software Version Number: R2.0

Reason for Recall

Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient.

Details

Recalling Firm
Philips North America Llc
Units Affected
338
Distribution
Global Distribution. US Nationwide.
Location
Cambridge, MA

Frequently Asked Questions

What product was recalled?
Philips Allura (Model Numbers 722010, 722012, 722026, 722027, 722028, 722029, 722035) and Philips Azurion (Model Numbers 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228) running StentBoost Live (Model Number 459801781661) Software Version Number: R2.0. Recalled by Philips North America Llc. Units affected: 338.
Why was this product recalled?
Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 13, 2022. Severity: Moderate. Recall number: Z-0874-2022.