PlainRecalls
FDA Devices Low Class III Ongoing

MM-TSM (Touch Screen Module/Secondary Controller/Bedside Controller), Part Numbers: 300002018101, 300002018102, used with IntraSight Integrated (Part Number: 797403) IVUS systems

Reported: April 20, 2022 Initiated: February 3, 2022 #Z-0917-2022

Product Description

MM-TSM (Touch Screen Module/Secondary Controller/Bedside Controller), Part Numbers: 300002018101, 300002018102, used with IntraSight Integrated (Part Number: 797403) IVUS systems

Reason for Recall

Multi-Modality Touch Screen Modules (MM-TSM), used with integrated IVUS systems, may crack, or become damaged, which may affect their usability. They could become non-functional. Therefore, bottom protective beams will be replaced with full protective metal brackets.

Details

Recalling Firm
Volcano Corp
Units Affected
1358
Distribution
US: AZ, IL, NV, MI, FL, NC, AK, PA, WI, MN, CA, AR, NJ, TX, OR, NM, SD, CO, KY, LA, DE, MA, MT, MO, VA, AL, NY, WV, SC, IN, GA, IA, OH, UT, NH, MS, TN, OK, ID, MD, HI, RI, WY, DC, WA, KS, NE, ND, CT, VT. OUS: SK, IT, DK, GB, DE, SE, IN, KW, QA, PL, NL, CL, JP, BE, AU, FR, PF, PT, NZ, CA, ES, PK, CY, AT, IL, IE, CH, GR, SA, PS, LB, HK, IS, TH, AR, OM, AE, NO, ID, RO, RS, MM, ZA, SG
Location
San Diego, CA

Frequently Asked Questions

What product was recalled?
MM-TSM (Touch Screen Module/Secondary Controller/Bedside Controller), Part Numbers: 300002018101, 300002018102, used with IntraSight Integrated (Part Number: 797403) IVUS systems. Recalled by Volcano Corp. Units affected: 1358.
Why was this product recalled?
Multi-Modality Touch Screen Modules (MM-TSM), used with integrated IVUS systems, may crack, or become damaged, which may affect their usability. They could become non-functional. Therefore, bottom protective beams will be replaced with full protective metal brackets.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 20, 2022. Severity: Low. Recall number: Z-0917-2022.