FDA Devices Moderate Class II Terminated

Integra CUSA excel. CEM Nosecone for CUSA EXcel System. The CEM Nosecone is intended for use CUSA Excel/CUSA Excel + Ultrasonic Surgical Aspiration System, CUSA handpiece, and Covidien Force FX Electrosurgical Generator. Used in surgical procedures where combined ultrasonic dissection and electrosurgical coagulation is desired, either simultaneously or independently.

Reported: March 20, 2013 Initiated: January 7, 2013 #Z-0920-2013

Product Description

Integra CUSA excel. CEM Nosecone for CUSA EXcel System. The CEM Nosecone is intended for use CUSA Excel/CUSA Excel + Ultrasonic Surgical Aspiration System, CUSA handpiece, and Covidien Force FX Electrosurgical Generator. Used in surgical procedures where combined ultrasonic dissection and electrosurgical coagulation is desired, either simultaneously or independently.

Reason for Recall

There is a potential for erosion of the CUSA EXcel tip used with the Nosecone. Specifically, the potential for tip erosion occurs when using a CUSA CEM Nosecone if the side of the tip, rather than the annulus of the tip, is in contact with the tissue. This erosion can potentially lead to tip breakage. There were no reports of a patient injury associated with these complaints.

Details

Units Affected
USA 6,190, EU 22,972 and ROW 288 units
Distribution
Worldwide Distribution - USA and the countries of Austria, Australia, Bangladesh, Belgium, Bulgaria, Brazil, Switzerland, Canada, China, Costa Rica, Czech Republic, Germany, Dominican Republic, Egypt, Emirates, Spain, Finland, France, Great Britian, Hong Kong, Hungary, Indonesia, Ireland, Israel, India, Italy, Japan, Korea (South), Kuwait, Lebanon, Sri Lanka, Latvia, Macedonia, Mexico, Malaysia, Netherlands, Philippines, Pakistan, Poland, Portugal, Qatar, Serbia, Saudi Arabia, Singapore, Slovenia, Thailand, Tunisia, Taiwan, United Arab, Viet Nam and South Africa.
Location
Plainsboro, NJ

Frequently Asked Questions

What product was recalled?
Integra CUSA excel. CEM Nosecone for CUSA EXcel System. The CEM Nosecone is intended for use CUSA Excel/CUSA Excel + Ultrasonic Surgical Aspiration System, CUSA handpiece, and Covidien Force FX Electrosurgical Generator. Used in surgical procedures where combined ultrasonic dissection and electrosurgical coagulation is desired, either simultaneously or independently.. Recalled by Integra LifeSciences Corp.. Units affected: USA 6,190, EU 22,972 and ROW 288 units.
Why was this product recalled?
There is a potential for erosion of the CUSA EXcel tip used with the Nosecone. Specifically, the potential for tip erosion occurs when using a CUSA CEM Nosecone if the side of the tip, rather than the annulus of the tip, is in contact with the tissue. This erosion can potentially lead to tip breakage. There were no reports of a patient injury associated with these complaints.
Which agency issued this recall?
This recall was issued by the FDA Devices on March 20, 2013. Severity: Moderate. Recall number: Z-0920-2013.