Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.

Reported: March 20, 2013 Initiated: December 13, 2012 #Z-0939-2013

Product Description

Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.

Reason for Recall

The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.

Details

Recalling Firm
Draeger Medical, Inc.
Units Affected
260
Distribution
Distributed in the states of CA, IL, and MA.
Location
Telford, PA

Frequently Asked Questions

What product was recalled?
Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.. Recalled by Draeger Medical, Inc.. Units affected: 260.
Why was this product recalled?
The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.
Which agency issued this recall?
This recall was issued by the FDA Devices on March 20, 2013. Severity: Moderate. Recall number: Z-0939-2013.