PlainRecalls
FDA Devices Moderate Class II Terminated

PreludeEASE Hydrophilic Sheath Introducer Dilator The Merit PreludeEASE Hydrophilic Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries, including but not limited to the radial artery, while maintaining hemostasis for a variety of diagnostic and therapeutic procedures. The access needle with inner metal needle and outer plastic cannula is used to gain access to the vein or artery for placement of guide wires

Reported: January 11, 2017 Initiated: December 1, 2016 #Z-0949-2017

Product Description

PreludeEASE Hydrophilic Sheath Introducer Dilator The Merit PreludeEASE Hydrophilic Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries, including but not limited to the radial artery, while maintaining hemostasis for a variety of diagnostic and therapeutic procedures. The access needle with inner metal needle and outer plastic cannula is used to gain access to the vein or artery for placement of guide wires

Reason for Recall

Recalling dilators included with the 6F PreludeEASE Hydrophilic Sheath Introducer because the dilator tips were manufactured with an undersized inner diameter and therefore will not allow passage of an 0.021 (0.53 mm) guide wire.

Details

Units Affected
335
Distribution
Nationwide Distribution
Location
South Jordan, UT

Frequently Asked Questions

What product was recalled?
PreludeEASE Hydrophilic Sheath Introducer Dilator The Merit PreludeEASE Hydrophilic Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries, including but not limited to the radial artery, while maintaining hemostasis for a variety of diagnostic and therapeutic procedures. The access needle with inner metal needle and outer plastic cannula is used to gain access to the vein or artery for placement of guide wires. Recalled by Merit Medical Systems, Inc.. Units affected: 335.
Why was this product recalled?
Recalling dilators included with the 6F PreludeEASE Hydrophilic Sheath Introducer because the dilator tips were manufactured with an undersized inner diameter and therefore will not allow passage of an 0.021 (0.53 mm) guide wire.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 11, 2017. Severity: Moderate. Recall number: Z-0949-2017.