FDA Devices Critical Class I Ongoing

VDR4 Phasitron Breathing Circuit Models that contain the venturi component: -A50094 D 5PK, Phasitron Kit, VDR, Single Patient, 5pk -A51001, VDR4 Control Unit Tester -PRT 991, VDR W/SWIVEL T SINGLE PATIENT PHASITRON -PRT 992, VDR4 HUMIDIFICATION ADAPTER KIT, CASE OF 10 -PRT 993, VDR4 Humidification Kit with Cross Tee, Case of 10 The Phasitron breathing circuit kit is intended to be used for continuous ventilation intended for controlled ventilation of patients

Reported: January 29, 2025 Initiated: December 23, 2024 #Z-0960-2025

Product Description

VDR4 Phasitron Breathing Circuit Models that contain the venturi component: -A50094 D 5PK, Phasitron Kit, VDR, Single Patient, 5pk -A51001, VDR4 Control Unit Tester -PRT 991, VDR W/SWIVEL T SINGLE PATIENT PHASITRON -PRT 992, VDR4 HUMIDIFICATION ADAPTER KIT, CASE OF 10 -PRT 993, VDR4 Humidification Kit with Cross Tee, Case of 10 The Phasitron breathing circuit kit is intended to be used for continuous ventilation intended for controlled ventilation of patients

Reason for Recall

Due to internal venturi component getting transiently stuck in the forward position within the breathing circuit and may cause a reduction of amplitude and volume of ventilation.

Details

Units Affected
4,727 packs/cases
Distribution
Worldwide - U.S. Nationwide distribution in the states of AZ, FL, IL, MS, MO, NE, NJ, NY, NC, RI, TX, UT, and WV. The countries of Belgium, Canada, France, Hong Kong, Netherlands, Russia, Switzerland and Turkey.
Location
Sandpoint, ID

Frequently Asked Questions

What product was recalled?
VDR4 Phasitron Breathing Circuit Models that contain the venturi component: -A50094 D 5PK, Phasitron Kit, VDR, Single Patient, 5pk -A51001, VDR4 Control Unit Tester -PRT 991, VDR W/SWIVEL T SINGLE PATIENT PHASITRON -PRT 992, VDR4 HUMIDIFICATION ADAPTER KIT, CASE OF 10 -PRT 993, VDR4 Humidification Kit with Cross Tee, Case of 10 The Phasitron breathing circuit kit is intended to be used for continuous ventilation intended for controlled ventilation of patients. Recalled by Percussionaire Corporation. Units affected: 4,727 packs/cases.
Why was this product recalled?
Due to internal venturi component getting transiently stuck in the forward position within the breathing circuit and may cause a reduction of amplitude and volume of ventilation.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 29, 2025. Severity: Critical. Recall number: Z-0960-2025.