BD Vacutainer¿ K2 EDTA (K2E) 3.6mg Blood Collection Tube (13 x 75 mm x 2.0 mL). Catalog Number: 367841. Used for whole blood hematology determinations, immunohematology testing and blood donor screening.

Reported: March 21, 2018 Initiated: August 29, 2017 #Z-0968-2018

Product Description

BD Vacutainer¿ K2 EDTA (K2E) 3.6mg Blood Collection Tube (13 x 75 mm x 2.0 mL). Catalog Number: 367841. Used for whole blood hematology determinations, immunohematology testing and blood donor screening.

Reason for Recall

A limited portion of the lot was manufactured with less than the required amount of K2EDTA additive, an anticoagulant, which may cause erroneous results that could lead to specimen recollection and potential delay of treatment or misdiagnosis/mismanagement of treatment and serious complications, such as failure to detect thrombocytosis or postponing surgery.

Details

Units Affected
8,494,000
Distribution
Nationwide Distribution
Location
Franklin Lakes, NJ

Frequently Asked Questions

What product was recalled?
BD Vacutainer¿ K2 EDTA (K2E) 3.6mg Blood Collection Tube (13 x 75 mm x 2.0 mL). Catalog Number: 367841. Used for whole blood hematology determinations, immunohematology testing and blood donor screening.. Recalled by Becton Dickinson & Company. Units affected: 8,494,000.
Why was this product recalled?
A limited portion of the lot was manufactured with less than the required amount of K2EDTA additive, an anticoagulant, which may cause erroneous results that could lead to specimen recollection and potential delay of treatment or misdiagnosis/mismanagement of treatment and serious complications, such as failure to detect thrombocytosis or postponing surgery.
Which agency issued this recall?
This recall was issued by the FDA Devices on March 21, 2018. Severity: Moderate. Recall number: Z-0968-2018.