MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

Reported: January 21, 2026 Initiated: December 10, 2025 #Z-1024-2026

Product Description

MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

Reason for Recall

Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.

Details

Recalling Firm
REPLIGEN CORPORATION
Units Affected
33
Distribution
U.S.
Location
Marlborough, MA

Frequently Asked Questions

What product was recalled?
MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496. Recalled by REPLIGEN CORPORATION. Units affected: 33.
Why was this product recalled?
Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.
Which agency issued this recall?
This recall was issued by the FDA Devices on January 21, 2026. Severity: Moderate. Recall number: Z-1024-2026.