PlainRecalls
FDA Devices Moderate Class II Ongoing

MEDLINE UNITE FOOT & ANKLE Screw, Non-Locking, Mini T8 (2.0x8mm), REF MMSN2008

Reported: February 8, 2023 Initiated: December 22, 2022 #Z-1045-2023

Product Description

MEDLINE UNITE FOOT & ANKLE Screw, Non-Locking, Mini T8 (2.0x8mm), REF MMSN2008

Reason for Recall

The recall is due to observed intra-operative screw failure. When pre-drilled with ¿1.3mm drill bit for ¿2.0mm screws and ¿1.6mm drill bit for ¿2.4mm we have observed screw failure due to excessive torque.

Details

Units Affected
420 units
Distribution
US Nationwide distribution.
Location
Northfield, IL

Frequently Asked Questions

What product was recalled?
MEDLINE UNITE FOOT & ANKLE Screw, Non-Locking, Mini T8 (2.0x8mm), REF MMSN2008. Recalled by MEDLINE INDUSTRIES, LP - Northfield. Units affected: 420 units.
Why was this product recalled?
The recall is due to observed intra-operative screw failure. When pre-drilled with ¿1.3mm drill bit for ¿2.0mm screws and ¿1.6mm drill bit for ¿2.4mm we have observed screw failure due to excessive torque.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 8, 2023. Severity: Moderate. Recall number: Z-1045-2023.