Siemens Dimension(R) EXL(TM) LM STM (Siemens Material Number 106393377). Intended to measure a variety of analytes in human body fluids.
Reported: April 10, 2013 Initiated: February 27, 2013 #Z-1047-2013
Product Description
Siemens Dimension(R) EXL(TM) LM STM (Siemens Material Number 106393377). Intended to measure a variety of analytes in human body fluids.
Reason for Recall
Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these errors are due to the current limit set for System Check LOCI results. The errors have no impact on results, but they may cause customer inconvenience. Siemens h
Details
- Recalling Firm
- Siemens Healthcare Diagnostics, Inc.
- Units Affected
- 2207 total
- Distribution
- Worldwide Distribution.
- Location
- Newark, DE
Frequently Asked Questions
What product was recalled? ▼
Siemens Dimension(R) EXL(TM) LM STM (Siemens Material Number 106393377). Intended to measure a variety of analytes in human body fluids.. Recalled by Siemens Healthcare Diagnostics, Inc.. Units affected: 2207 total.
Why was this product recalled? ▼
Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these errors are due to the current limit set for System Check LOCI results. The errors have no impact on results, but they may cause customer inconvenience. Siemens h
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on April 10, 2013. Severity: Low. Recall number: Z-1047-2013.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11