Allura Xper FD10; Model Number: 722003;
Reported: January 21, 2026 Initiated: December 15, 2025 #Z-1064-2026
Product Description
Allura Xper FD10; Model Number: 722003;
Reason for Recall
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Details
- Recalling Firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Units Affected
- 93 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen.
- Location
- Best
Frequently Asked Questions
What product was recalled? ▼
Allura Xper FD10; Model Number: 722003;. Recalled by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. Units affected: 93 units.
Why was this product recalled? ▼
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on January 21, 2026. Severity: Moderate. Recall number: Z-1064-2026.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11