PlainRecalls
FDA Devices Moderate Class II Terminated

STA - Neoplastine¿ Cl Plus ¿ (ref. 00606) Product Usage: Manual or automated determination of the prothrombin time (PT).

Reported: March 21, 2018 Initiated: October 25, 2017 #Z-1069-2018

Product Description

STA - Neoplastine¿ Cl Plus ¿ (ref. 00606) Product Usage: Manual or automated determination of the prothrombin time (PT).

Reason for Recall

Confirmed defect of homogeneity where some reagent vials will give prolonged Prothrombin Time (decreased PT%).

Details

Recalling Firm
Diagnostica Stago, Inc.
Distribution
Nationally
Location
Parsippany, NJ

Frequently Asked Questions

What product was recalled?
STA - Neoplastine¿ Cl Plus ¿ (ref. 00606) Product Usage: Manual or automated determination of the prothrombin time (PT).. Recalled by Diagnostica Stago, Inc..
Why was this product recalled?
Confirmed defect of homogeneity where some reagent vials will give prolonged Prothrombin Time (decreased PT%).
Which agency issued this recall?
This recall was issued by the FDA Devices on March 21, 2018. Severity: Moderate. Recall number: Z-1069-2018.