PlainRecalls
FDA Devices Critical Class I Ongoing

Intact Vascular Tack Endovascular System, to treat vascular dissections with Tack implant(s) following angioplasty: (4F,1.5-4.5mm), 150cm, REF: 154150041; Tack, 4F Gen 1.5, 150cm CE, REF: 154150042; Tack, 4F Gen 1.5, 90cm CE, REF: 154090042 (6F, 3.5 - 6.0mm), 135cm, REF: 156135061; (6F, 4.0 - 8.0mm), 135cm, REF: 206135061; Tack, 6F Gen 2.0, 135cm CE, REF: 206135062; Tack, 6F Gen 1.5, 135cm CE, REF: 156135062; Tack, 6F Gen 2.0, 80cm CE, REF: 206080062; Tack, 6F Gen 1.5, 80cm CE, REF: 15608006

Reported: February 26, 2025 Initiated: January 10, 2025 #Z-1070-2025

Product Description

Intact Vascular Tack Endovascular System, to treat vascular dissections with Tack implant(s) following angioplasty: (4F,1.5-4.5mm), 150cm, REF: 154150041; Tack, 4F Gen 1.5, 150cm CE, REF: 154150042; Tack, 4F Gen 1.5, 90cm CE, REF: 154090042 (6F, 3.5 - 6.0mm), 135cm, REF: 156135061; (6F, 4.0 - 8.0mm), 135cm, REF: 206135061; Tack, 6F Gen 2.0, 135cm CE, REF: 206135062; Tack, 6F Gen 1.5, 135cm CE, REF: 156135062; Tack, 6F Gen 2.0, 80cm CE, REF: 206080062; Tack, 6F Gen 1.5, 80cm CE, REF: 156080062

Reason for Recall

Use of Tack Endovascular system, designed to treat acute dissections of inner wall or lining of an artery by tacking damaged tissue to inner luminal surface, could result in problems with deployment and stability of device in vessel after deployment, which may cause failure to resolve dissection, migration of implant, bailout stenting, reintervention, unintended removal of tack devices, and ischemia.

Details

Recalling Firm
Spectranetics Corporation
Units Affected
2,939
Distribution
Distribution US nationwide and countries of: NL, DE, NL, AU, IT, CH, GB, and AT
Location
Colorado Springs, CO

Frequently Asked Questions

What product was recalled?
Intact Vascular Tack Endovascular System, to treat vascular dissections with Tack implant(s) following angioplasty: (4F,1.5-4.5mm), 150cm, REF: 154150041; Tack, 4F Gen 1.5, 150cm CE, REF: 154150042; Tack, 4F Gen 1.5, 90cm CE, REF: 154090042 (6F, 3.5 - 6.0mm), 135cm, REF: 156135061; (6F, 4.0 - 8.0mm), 135cm, REF: 206135061; Tack, 6F Gen 2.0, 135cm CE, REF: 206135062; Tack, 6F Gen 1.5, 135cm CE, REF: 156135062; Tack, 6F Gen 2.0, 80cm CE, REF: 206080062; Tack, 6F Gen 1.5, 80cm CE, REF: 156080062. Recalled by Spectranetics Corporation. Units affected: 2,939.
Why was this product recalled?
Use of Tack Endovascular system, designed to treat acute dissections of inner wall or lining of an artery by tacking damaged tissue to inner luminal surface, could result in problems with deployment and stability of device in vessel after deployment, which may cause failure to resolve dissection, migration of implant, bailout stenting, reintervention, unintended removal of tack devices, and ischemia.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 26, 2025. Severity: Critical. Recall number: Z-1070-2025.