Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet), REF 8184.
Reported: February 21, 2024 Initiated: January 8, 2024 #Z-1075-2024
Product Description
Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet), REF 8184.
Reason for Recall
The negative control was contaminated with one of the positive control pathogens (Norovirus). This would result in users getting a positive call for Norovirus when running the negative control.
Details
- Recalling Firm
- Microbiologics Inc
- Units Affected
- 580 packages
- Distribution
- US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MN, MO, MS, MT, NE, NH, NJ, NY, OH, PA, RI, TN, TX, VA, WA, WI, WV, and WY. There was government distribution and no military distribution. Foreign distribution was made to France.
- Location
- Saint Cloud, MN
Frequently Asked Questions
What product was recalled? ▼
Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet), REF 8184.. Recalled by Microbiologics Inc. Units affected: 580 packages.
Why was this product recalled? ▼
The negative control was contaminated with one of the positive control pathogens (Norovirus). This would result in users getting a positive call for Norovirus when running the negative control.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on February 21, 2024. Severity: Moderate. Recall number: Z-1075-2024.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11