FDA Devices Moderate Class II Terminated

Crosslock DVR ePAK -Sterile, Ref 81180050, are complete surgical sets including plates, screws, and instruments for orthopedic fixation The Cross lock DVR ePAK Volar Rim Plates system is intended for fixation of fractures, mal unions, and osteotomies involving the distal radius. The DVR Wrist ePAK system is intended stabilization and fixation of small bone fragments in fresh fractures, revision procedures, joint fusion and reconstructions of small bones of the hand, foot, radius, ulna, ankle,

Reported: March 16, 2016 Initiated: January 29, 2016 #Z-1082-2016

Product Description

Crosslock DVR ePAK -Sterile, Ref 81180050, are complete surgical sets including plates, screws, and instruments for orthopedic fixation The Cross lock DVR ePAK Volar Rim Plates system is intended for fixation of fractures, mal unions, and osteotomies involving the distal radius. The DVR Wrist ePAK system is intended stabilization and fixation of small bone fragments in fresh fractures, revision procedures, joint fusion and reconstructions of small bones of the hand, foot, radius, ulna, ankle, humerus, scapula and pelvis, particularly in osteopenic bone.

Reason for Recall

Components of the DVR ePAKs can oxidize during shipment and storage prior to use. A delay in surgery, less than 30 minutes, may occur if the surgical staff proceeds to re-sterilize the product. If oxidative material would flake off a component contained within the ePAK and enter the surgical field/wound an inflammatory reaction may occur. The inflammatory reaction may result in a foreign body reaction and/or metal hypersensitivity. Oxidized components have failed cytotoxicity testing.

Details

Recalling Firm
Biomet, Inc.
Units Affected
21,268
Distribution
Worldwide distribution: US (Nationwide) and countries of: Panama; Korea; Japan; Netherlands; Ecuador; Chile; Canada; and Australia.
Location
Warsaw, IN

Frequently Asked Questions

What product was recalled?
Crosslock DVR ePAK -Sterile, Ref 81180050, are complete surgical sets including plates, screws, and instruments for orthopedic fixation The Cross lock DVR ePAK Volar Rim Plates system is intended for fixation of fractures, mal unions, and osteotomies involving the distal radius. The DVR Wrist ePAK system is intended stabilization and fixation of small bone fragments in fresh fractures, revision procedures, joint fusion and reconstructions of small bones of the hand, foot, radius, ulna, ankle, humerus, scapula and pelvis, particularly in osteopenic bone.. Recalled by Biomet, Inc.. Units affected: 21,268.
Why was this product recalled?
Components of the DVR ePAKs can oxidize during shipment and storage prior to use. A delay in surgery, less than 30 minutes, may occur if the surgical staff proceeds to re-sterilize the product. If oxidative material would flake off a component contained within the ePAK and enter the surgical field/wound an inflammatory reaction may occur. The inflammatory reaction may result in a foreign body reaction and/or metal hypersensitivity. Oxidized components have failed cytotoxicity testing.
Which agency issued this recall?
This recall was issued by the FDA Devices on March 16, 2016. Severity: Moderate. Recall number: Z-1082-2016.