9180 Electrolyte Analyzer - Product Usage: intended to be used for the measurement of sodium, potassium, chloride, ionized calcium and lithium in the whole blood, serum or plasma, urine, dialysate solutions, or QC materials. catalog number 03157334001 . in vitro diagnostic use.

Reported: February 17, 2021 Initiated: January 6, 2021 #Z-1098-2021

Product Description

9180 Electrolyte Analyzer - Product Usage: intended to be used for the measurement of sodium, potassium, chloride, ionized calcium and lithium in the whole blood, serum or plasma, urine, dialysate solutions, or QC materials. catalog number 03157334001 . in vitro diagnostic use.

Reason for Recall

May display Calcium results on the screen with wrong arrow direction which may lead to misinterpretation of results and incorrect medical decision. If the unit for iCa++ is set to mg/dL (configured as Service Code MGL) and a measured iCa++ value is lower than the normal range, an upward arrow is shown on the display instead of a downward arrow. The display is set per default to mmol/L; therefore, the arrow indicating an alarm reflects normal ranges of iCa++ in mmol/L. If mg/dL is chosen, the arrow warnings on screen still reflect the ranges from mmol/L, rather than mg/dL. The numeric result is displayed correctly.

Details

Units Affected
1,214 units
Distribution
US Nationwide distribution.
Location
Indianapolis, IN

Frequently Asked Questions

What product was recalled?
9180 Electrolyte Analyzer - Product Usage: intended to be used for the measurement of sodium, potassium, chloride, ionized calcium and lithium in the whole blood, serum or plasma, urine, dialysate solutions, or QC materials. catalog number 03157334001 . in vitro diagnostic use.. Recalled by Roche Diagnostics Operations, Inc.. Units affected: 1,214 units.
Why was this product recalled?
May display Calcium results on the screen with wrong arrow direction which may lead to misinterpretation of results and incorrect medical decision. If the unit for iCa++ is set to mg/dL (configured as Service Code MGL) and a measured iCa++ value is lower than the normal range, an upward arrow is shown on the display instead of a downward arrow. The display is set per default to mmol/L; therefore, the arrow indicating an alarm reflects normal ranges of iCa++ in mmol/L. If mg/dL is chosen, the arrow warnings on screen still reflect the ranges from mmol/L, rather than mg/dL. The numeric result is displayed correctly.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 17, 2021. Severity: Moderate. Recall number: Z-1098-2021.