PlainRecalls
FDA Devices Moderate Class II Terminated

SORIN Heater-Cooler System 3T, REF 16-02-85 Intended for use with a Stockert/Sorin S3/S5/C5 heart-lung machine and/or any other heart-lung machine featuring a separate temperature control for extracorporeal perfusion of durations of up to 6 hours.

Reported: April 17, 2019 Initiated: September 14, 2018 #Z-1099-2019

Product Description

SORIN Heater-Cooler System 3T, REF 16-02-85 Intended for use with a Stockert/Sorin S3/S5/C5 heart-lung machine and/or any other heart-lung machine featuring a separate temperature control for extracorporeal perfusion of durations of up to 6 hours.

Reason for Recall

For six units, the post-deep cleaning (DC) water sample tested positive for nontuberculous mycobacteria (NTM).

Details

Recalling Firm
LivaNova USA Inc.
Units Affected
6 devices
Distribution
US Distribution to states of: IN, MO, and OH.
Location
Arvada, CO

Frequently Asked Questions

What product was recalled?
SORIN Heater-Cooler System 3T, REF 16-02-85 Intended for use with a Stockert/Sorin S3/S5/C5 heart-lung machine and/or any other heart-lung machine featuring a separate temperature control for extracorporeal perfusion of durations of up to 6 hours.. Recalled by LivaNova USA Inc.. Units affected: 6 devices.
Why was this product recalled?
For six units, the post-deep cleaning (DC) water sample tested positive for nontuberculous mycobacteria (NTM).
Which agency issued this recall?
This recall was issued by the FDA Devices on April 17, 2019. Severity: Moderate. Recall number: Z-1099-2019.