Malecot Nephrostomy Catheter Set, Product Number: M0064101000, GTIN: 08714729005261 The Malecot Nephrostomy Catheter Set is intended to establish percutaneous nephrostomy drainage and is indicated for the following: Diagnostic Indications: " Antegrade pyelography " Pressure/perfusion study (Whitaker test) Therapeutic Indications: " Nephrostomy catheter drainage " Perfusion chemolysis of renal stones " Percutaneous nephrolithotomy " Percutaneous resection and coagulation of urothelial tumors

Reported: March 28, 2018 Initiated: December 12, 2017 #Z-1100-2018

Product Description

Malecot Nephrostomy Catheter Set, Product Number: M0064101000, GTIN: 08714729005261 The Malecot Nephrostomy Catheter Set is intended to establish percutaneous nephrostomy drainage and is indicated for the following: Diagnostic Indications: " Antegrade pyelography " Pressure/perfusion study (Whitaker test) Therapeutic Indications: " Nephrostomy catheter drainage " Perfusion chemolysis of renal stones " Percutaneous nephrolithotomy " Percutaneous resection and coagulation of urothelial tumors

Reason for Recall

Certain Malecot Nephrostomy Catheters are breaking at the mid-shaft bond during use.

Details

Units Affected
1480
Distribution
Worldwide Distribution: US (nationwide) in the following: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KA, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, including the District of Columbia and Puerto Rico; and countries of: Argentina, Bolivia, Brazil, Canada, Chile, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Mexico, Panama, Peru, Venezuela, Australia, Bangladesh, Botswana, China, Kuwait, Lebanon, Macau, Malaysia, Philippines, Saudi Arabia, Singapore, South Africa, South Korea, Turkey, United Arab Emirates, and Zimbabwe.
Location
Marlborough, MA

Frequently Asked Questions

What product was recalled?
Malecot Nephrostomy Catheter Set, Product Number: M0064101000, GTIN: 08714729005261 The Malecot Nephrostomy Catheter Set is intended to establish percutaneous nephrostomy drainage and is indicated for the following: Diagnostic Indications: " Antegrade pyelography " Pressure/perfusion study (Whitaker test) Therapeutic Indications: " Nephrostomy catheter drainage " Perfusion chemolysis of renal stones " Percutaneous nephrolithotomy " Percutaneous resection and coagulation of urothelial tumors. Recalled by Boston Scientific Corporation. Units affected: 1480.
Why was this product recalled?
Certain Malecot Nephrostomy Catheters are breaking at the mid-shaft bond during use.
Which agency issued this recall?
This recall was issued by the FDA Devices on March 28, 2018. Severity: Moderate. Recall number: Z-1100-2018.