Panaroma Central Station Product Usage: This device can view real time, store, print, graph, and trend patient clinical and demographic data. The device can also set independent alarm limits for data send by the bedside monitor.
Product Description
Panaroma Central Station Product Usage: This device can view real time, store, print, graph, and trend patient clinical and demographic data. The device can also set independent alarm limits for data send by the bedside monitor.
Reason for Recall
Mindray has identified two issues with the Panorama Central Station. Full Disclosure data may lag behind the Panorama system time clock when the Panorama Central Station is in use with Mindray¿s V-Series Monitor. Additionally, event indicators lag behind the Panorama system time clock when the Panorama Central Station is in use with a Panorama Telepack.
Details
- Recalling Firm
- Mindray DS USA, Inc. dba Mindray North America
- Units Affected
- 732 units in the US
- Distribution
- Worldwide Distribution - USA Nationwide Distribution
- Location
- Mahwah, NJ
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11