Smith & Nephew RENASYS-AB Abdominal Dressing Kit with Soft Port REF #'s: 66800980 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)

Reported: March 19, 2014 Initiated: January 9, 2014 #Z-1162-2014

Product Description

Smith & Nephew RENASYS-AB Abdominal Dressing Kit with Soft Port REF #'s: 66800980 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)

Reason for Recall

Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo

Details

Recalling Firm
Smith & Nephew Inc.
Units Affected
28,250 individual units
Distribution
Worldwide Distribution-USA (nationwide) including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, KY, LA, MA, MS, MT, NC, NJ, NV, NY, OH, OR, PA, TN, TX, VA, WA, WI, WV, HI and the countries Dubai, Curacao, Argentina, Australia, Chile, Germany, France, England, Hong Kong, Kuwait, Lebanon, Libya, Sri Lanka, Malaysia, New Zealand, Manila, Saudi Arabia, South Africa, Singapore, Turkey, Taipei and Vietnam.
Location
Saint Petersburg, FL

Frequently Asked Questions

What product was recalled?
Smith & Nephew RENASYS-AB Abdominal Dressing Kit with Soft Port REF #'s: 66800980 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps). Recalled by Smith & Nephew Inc.. Units affected: 28,250 individual units.
Why was this product recalled?
Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo
Which agency issued this recall?
This recall was issued by the FDA Devices on March 19, 2014. Severity: Moderate. Recall number: Z-1162-2014.