Sofia 2 Analyzer Catalog Number 20299 Sofia 2 is a bench top analyzer intended to be used with Cassette-based immunofluorescent in vitro diagnostic assays . Sofia 2 is intended for professional and laboratory use.

Reported: April 4, 2018 Initiated: December 15, 2017 #Z-1183-2018

Product Description

Sofia 2 Analyzer Catalog Number 20299 Sofia 2 is a bench top analyzer intended to be used with Cassette-based immunofluorescent in vitro diagnostic assays . Sofia 2 is intended for professional and laboratory use.

Reason for Recall

The calibration for the affected batch of analyzers was incorrectly set during the manufacturing process which may lead to a failure when calibration is first performed. There may also be e remote possibility that incorrect qualitative test results for patient samples near the assay cutoff may be reported.

Details

Recalling Firm
Quidel Corporation
Units Affected
482 units
Distribution
US Distribution
Location
San Diego, CA

Frequently Asked Questions

What product was recalled?
Sofia 2 Analyzer Catalog Number 20299 Sofia 2 is a bench top analyzer intended to be used with Cassette-based immunofluorescent in vitro diagnostic assays . Sofia 2 is intended for professional and laboratory use.. Recalled by Quidel Corporation. Units affected: 482 units.
Why was this product recalled?
The calibration for the affected batch of analyzers was incorrectly set during the manufacturing process which may lead to a failure when calibration is first performed. There may also be e remote possibility that incorrect qualitative test results for patient samples near the assay cutoff may be reported.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 4, 2018. Severity: Moderate. Recall number: Z-1183-2018.