Chest Drainage Units and Accessories: Product Code Description 1. 8888571562 Cardinal Health Sentinel Seal CDU 2. 8888571513 Cardinal Health Sentinel Seal CDU, Dual Drain 3. 8888571489 Cardinal Health Sentinel Seal CDU, with Easy Change Connector 4. 8888571370 Cardinal Health Altitude CDU, Dry Suction, with Easy Change Connector 5. 8888571299 Cardinal Health Aqua-Seal CDU, Wet Suction 6. 8888571406 Cardinal Health Aqua-Seal CDU, Wet Suction, Dual Drain 7. 8888571315 Cardinal Health
Product Description
Chest Drainage Units and Accessories: Product Code Description 1. 8888571562 Cardinal Health Sentinel Seal CDU 2. 8888571513 Cardinal Health Sentinel Seal CDU, Dual Drain 3. 8888571489 Cardinal Health Sentinel Seal CDU, with Easy Change Connector 4. 8888571370 Cardinal Health Altitude CDU, Dry Suction, with Easy Change Connector 5. 8888571299 Cardinal Health Aqua-Seal CDU, Wet Suction 6. 8888571406 Cardinal Health Aqua-Seal CDU, Wet Suction, Dual Drain 7. 8888571315 Cardinal Health Aqua-Seal CDU, Wet Suction, with Easy Change Connector 8. 8884713100 Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL 9. 1814713105 OUS¿Only - Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL (Bottle only) 10. 1180571570 OUS¿Only - Cardinal Health Thora-Seal CDU, Two Chamber System, 2600 mL 11. 8884713308 Cardinal Health Thora-Seal CDU, Three Chamber System, 2500 mL 12. 8884713900 Cardinal Health Thora-Seal Replacement Collection Chamber, 2500 mL
Reason for Recall
The firm is updating the IFU to clarify the intended patient population for the identified chest drainage units and accessories as adults, 18 years and older. When used on infants, the CDU may not demonstrate tidaling/bubbling, which may confound the clinician's assessment of device functionality and lead to inadequate/inappropriate treatment/therapy, delay to treatment/therapy, and prolonged hospitalization.
Details
- Recalling Firm
- Cardinal Health 200, LLC
- Units Affected
- 1,169,726 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Brazil, Canada, Chile, Colombia, Dominican Republic, and Mexico.
- Location
- Waukegan, IL
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11