FreeStyle Blood Glucose Monitors System. For in vitro diagnostic use only. Store at room temperature below 86 degrees F (30 degrees C). Use test strips only within the system operating temperature range as outlined in your Owner's Booklet. Meter made in China; Distributed by Abbott Diabetes Care, Alameda, CA The FreeStyle mete should only be used with FreeStyle Test Strips and FreeStyle Control Solution. Using other brands of test strips and control solutions with the FreeStyle Mete c
Product Description
FreeStyle Blood Glucose Monitors System. For in vitro diagnostic use only. Store at room temperature below 86 degrees F (30 degrees C). Use test strips only within the system operating temperature range as outlined in your Owner's Booklet. Meter made in China; Distributed by Abbott Diabetes Care, Alameda, CA The FreeStyle mete should only be used with FreeStyle Test Strips and FreeStyle Control Solution. Using other brands of test strips and control solutions with the FreeStyle Mete can product inaccurate results. The FreeStyle Glucose meter is incorporated into the OmniPod insulin Management system (manufactured by Insulet Corporation, Bedford, MA and is intended for subcutaneous delivery of insulin. Intended to monitor blood glucose from samples taken from the body. IVD use only,
Reason for Recall
Abbott Diabetes Care has identified through internal testing and investigation that all non-applied voltage legacy meters have the potential to produce out of range control solution results and/or erroneously low blood glucose results when used in conjunction with Free Style test strips lots within expiry.
Details
- Recalling Firm
- Abbott Diabetes Care, Inc.
- Units Affected
- not provided. Device has not been shipped in the last 5 years.
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: Austria, Canada, Germany, Israel, Netherlands, Norway, Sweden, Switzerland, United Kingdom.
- Location
- Alameda, CA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11