FDA Devices Low Class III Terminated

The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine Aminotransferase (ALTI and ALTI) and Aspartate Aminotransferase (AST) methods on the Dimension Vista System. For prescription use only.

Reported: February 22, 2017 Initiated: May 5, 2016 #Z-1216-2017

Product Description

The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine Aminotransferase (ALTI and ALTI) and Aspartate Aminotransferase (AST) methods on the Dimension Vista System. For prescription use only.

Reason for Recall

Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vista Enzyme 2 calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.

Details

Units Affected
13,147 units
Distribution
Distributed throughout the United States
Location
Newark, DE

Frequently Asked Questions

What product was recalled?
The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine Aminotransferase (ALTI and ALTI) and Aspartate Aminotransferase (AST) methods on the Dimension Vista System. For prescription use only.. Recalled by Siemens Healthcare Diagnostics, Inc.. Units affected: 13,147 units.
Why was this product recalled?
Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vista Enzyme 2 calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 22, 2017. Severity: Low. Recall number: Z-1216-2017.