CardioMEMS HF System PA Sensor and Delivery System, Model Number CM3100, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which physicians use to initiate or modify heart failure treatment.

Reported: March 15, 2023 Initiated: February 7, 2023 #Z-1225-2023

Product Description

CardioMEMS HF System PA Sensor and Delivery System, Model Number CM3100, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which physicians use to initiate or modify heart failure treatment.

Reason for Recall

Select CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of the intended radiofrequency has the potential to result in inaccurate readings or sensor signal acquisition difficulties.

Details

Recalling Firm
St. Jude Medical
Units Affected
317 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK, United Arab Emirates.
Location
Atlanta, GA

Frequently Asked Questions

What product was recalled?
CardioMEMS HF System PA Sensor and Delivery System, Model Number CM3100, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which physicians use to initiate or modify heart failure treatment.. Recalled by St. Jude Medical. Units affected: 317 units.
Why was this product recalled?
Select CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of the intended radiofrequency has the potential to result in inaccurate readings or sensor signal acquisition difficulties.
Which agency issued this recall?
This recall was issued by the FDA Devices on March 15, 2023. Severity: Moderate. Recall number: Z-1225-2023.