PlainRecalls
FDA Devices Moderate Class II Terminated

Stryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip, vented, Non-Reflective Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.

Reported: May 15, 2013 Initiated: March 27, 2013 #Z-1230-2013

Product Description

Stryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip, vented, Non-Reflective Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.

Reason for Recall

The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.

Details

Recalling Firm
Stryker Endoscopy
Units Affected
147,921 - all sizes
Distribution
Worldwide Distribution: USA (nationwide) and countries of: Mexico, Portugal, Latin America, India, Greece, Taiwan, Singapore, New Zealand, Phillipines, Netherlands, France, Australia, Canada, Japan, Scandinavia, Iberia, Brasil, Romania, South Africa, Chile, Korea, China.
Location
San Jose, CA

Frequently Asked Questions

What product was recalled?
Stryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip, vented, Non-Reflective Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.. Recalled by Stryker Endoscopy. Units affected: 147,921 - all sizes.
Why was this product recalled?
The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
Which agency issued this recall?
This recall was issued by the FDA Devices on May 15, 2013. Severity: Moderate. Recall number: Z-1230-2013.