PlainRecalls
FDA Devices Moderate Class II Terminated

Natural-Knee¿ N-K II DOMED ALL-POLY PAT, SZ0 N-K II DOMED ALL-POLY PAT, SZ1

Reported: March 26, 2014 Initiated: February 25, 2014 #Z-1249-2014

Product Description

Natural-Knee¿ N-K II DOMED ALL-POLY PAT, SZ0 N-K II DOMED ALL-POLY PAT, SZ1

Reason for Recall

A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.

Details

Recalling Firm
Zimmer, Inc.
Units Affected
3638 units
Distribution
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
Location
Warsaw, IN

Frequently Asked Questions

What product was recalled?
Natural-Knee¿ N-K II DOMED ALL-POLY PAT, SZ0 N-K II DOMED ALL-POLY PAT, SZ1. Recalled by Zimmer, Inc.. Units affected: 3638 units.
Why was this product recalled?
A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
Which agency issued this recall?
This recall was issued by the FDA Devices on March 26, 2014. Severity: Moderate. Recall number: Z-1249-2014.