Atellica CH Triglycerides (concentrated)-In vitro diagnostic use in the quantitative measurement of triglycerides in human serum and plasma (lithium heparin, potassium EDTA) using the Atellica CH Analyzer SMN: 11097591
Reported: September 6, 2023 Initiated: January 27, 2023 #Z-1270-2023
Product Description
Atellica CH Triglycerides (concentrated)-In vitro diagnostic use in the quantitative measurement of triglycerides in human serum and plasma (lithium heparin, potassium EDTA) using the Atellica CH Analyzer SMN: 11097591
Reason for Recall
Reagent Carryover from LDL Cholesterol Direct (DLDL), Total Protein II (TP), Triglycerides (Trig), and Triglycerides_2 (Trig_2) Impacting Magnesium (Mg) Results
Details
- Recalling Firm
- Siemens Healthcare Diagnostics, Inc.
- Units Affected
- 11824 units
- Distribution
- Nationwide Foreign: Country Albania Algeria Argentina Australia Austria Bahrain Bangladesh Belgium Brazil Bulgaria Burkina Faso Canada Chile China Colombia Croatia Cura¿ao Czech Republic Denmark Ecuador Egypt Finland France Germany Greece Guadeloupe Hong Kong Hungary India Ireland Israel Italy Japan Kuwait Libya Malaysia Mexico Netherlands New Zealand Norway Oman Pakistan Paraguay Peru Philippines Poland Portugal Qatar Republic Korea Romania Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan, Province of China Thailand Ukraine United Arab Emirates United Kingdom Great Britain Uruguay Vatikancity Vietnam
- Location
- Tarrytown, NY
Frequently Asked Questions
What product was recalled? ▼
Atellica CH Triglycerides (concentrated)-In vitro diagnostic use in the quantitative measurement of triglycerides in human serum and plasma (lithium heparin, potassium EDTA) using the Atellica CH Analyzer SMN: 11097591. Recalled by Siemens Healthcare Diagnostics, Inc.. Units affected: 11824 units.
Why was this product recalled? ▼
Reagent Carryover from LDL Cholesterol Direct (DLDL), Total Protein II (TP), Triglycerides (Trig), and Triglycerides_2 (Trig_2) Impacting Magnesium (Mg) Results
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on September 6, 2023. Severity: Moderate. Recall number: Z-1270-2023.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11