FDA Devices Moderate Class II Terminated

CIOS ALPHA (Image-intensified fluoroscope x-ray system) Product Usage: The Cios Alpha is a mobile X-ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric pat

Reported: April 6, 2016 Initiated: February 16, 2016 #Z-1281-2016

Product Description

CIOS ALPHA (Image-intensified fluoroscope x-ray system) Product Usage: The Cios Alpha is a mobile X-ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.

Reason for Recall

Siemens found a defective part in their production line which could increase risk of an electrical shock for service personnel when working inside the device during system maintenance. The risk of an electrical shock is only present when a service technician ignores several steps listed in the service manual as precautionary actions during device maintenance.

Details

Units Affected
4 units
Distribution
US Nationwide Distributed in the states of AR, CA, NE and UT
Location
Malvern, PA

Frequently Asked Questions

What product was recalled?
CIOS ALPHA (Image-intensified fluoroscope x-ray system) Product Usage: The Cios Alpha is a mobile X-ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.. Recalled by Siemens Medical Solutions USA, Inc. Units affected: 4 units.
Why was this product recalled?
Siemens found a defective part in their production line which could increase risk of an electrical shock for service personnel when working inside the device during system maintenance. The risk of an electrical shock is only present when a service technician ignores several steps listed in the service manual as precautionary actions during device maintenance.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 6, 2016. Severity: Moderate. Recall number: Z-1281-2016.