Clariti 1-day toric soft contact lenses, Power -1.00 -1.25/90

Reported: March 29, 2023 Initiated: February 14, 2023 #Z-1282-2023

Product Description

Clariti 1-day toric soft contact lenses, Power -1.00 -1.25/90

Reason for Recall

The lenses manufactured in the affected lot were made with a misaligned axis resulting in lenses with the incorrect power.

Details

Recalling Firm
CooperVision, Inc.
Units Affected
3150
Distribution
Worldwide - US Nationwide distribution in the states of FL, IL, NY, UT, and WI. The country of Canada.
Location
West Henrietta, NY

Frequently Asked Questions

What product was recalled?
Clariti 1-day toric soft contact lenses, Power -1.00 -1.25/90. Recalled by CooperVision, Inc.. Units affected: 3150.
Why was this product recalled?
The lenses manufactured in the affected lot were made with a misaligned axis resulting in lenses with the incorrect power.