FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)

Reported: February 11, 2026 Initiated: January 9, 2026 #Z-1299-2026

Product Description

FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)

Reason for Recall

It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.

Details

Units Affected
16 units
Distribution
US Nationwide distribution in the states of AK, CA, CT, FL, IL, NH, OH, WI.
Location
Lexington, MA

Frequently Asked Questions

What product was recalled?
FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3). Recalled by FUJIFILM Healthcare Americas Corporation. Units affected: 16 units.
Why was this product recalled?
It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.
Which agency issued this recall?
This recall was issued by the FDA Devices on February 11, 2026. Severity: Moderate. Recall number: Z-1299-2026.