GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed

Reported: March 19, 2025 Initiated: February 18, 2025 #Z-1301-2025

Product Description

GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed

Reason for Recall

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

Details

Recalling Firm
GE Medical Systems, LLC
Units Affected
1 unit
Distribution
Worldwide distribution.
Location
Waukesha, WI

Frequently Asked Questions

What product was recalled?
GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed. Recalled by GE Medical Systems, LLC. Units affected: 1 unit.
Why was this product recalled?
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.